Features
- 01 LCD display, touch control, easy to useÂ
- 02 It can be heated to 35 °C or cooled to-70 °C
- 03 Automatic detection, good repeatability, printable resultsÂ
- 04 Automatic tilt during detection, fully compliant with standards
- 05 The test method is optional, and it can be detected quickly or normally
- 06 It has fault diagnosis function and can prompt operation
- 07 Refrigeration speed and testing standards can be compiled according to customer requirements
- 08 High performance compressor refrigeration for long service life
- 09 Provide manual and automatic integrated test solutions, flexible and simple
working principle
Working principle ofTP526 Solidification point and Pour Point Tester
When the sample is packed in a specified test tube and cooled to the expected temperature, the push rod motor tilts the tilting rack to 45 degrees. After one minute, the high-precision light sensor detection technology is used to detect whether the liquid level condenses according to the reflection principle of light.
Technical indicators
|
Temperature control range |
35℃~-70℃ |
|
Solidification point range |
Light oil in the range of 5 ℃ ~-70 ℃; |
|
Display Mode: |
Color LCD touch screen Chinese display |
|
Resolution |
0.1 ℃ |
|
Temperature control accuracy |
±0.1℃ |
| Number of experimental holes |
1Â hole |
|
Temperature control method |
PIDÂ adjustment |
| Refrigeration system |
Two-stage refrigeration |
|
Test method |
standard mode, fast mode, manual mode; |
| Solidification point detection interval |
(1-5) ℃ adjustable, default 2℃; |
|
Tilt time |
(5-99) s adjustable, default 60 s; |
| Ambient temperature |
5~40℃ |
| Ambient humidity |
≤80%RH (no condensation) |
|
Power supply |
AC220V±10%    50Hz±10% |
| Power |
≤600W |
|
Overall dimensions |
500 mm×490 mm ×660 mm |
| Instrument weight |
55Â kg |
Ordering Guide
-
-
Accessories Guide
Freezing point test tube/pour point test tube/thermometer
-
-
Precautions
1. If the pour point meter is found to have obvious faults, the manufacturer should be contacted in time. Users are prohibited from opening it for repair.
2. If the pour point meter has no display after being turned on, check whether the fuse at the three-core power cord interface behind the instrument is intact. If the fuse is broken, the fuse should be replaced.
3. If the pour point meter fails, the power supply should be cut off immediately. After the cold tank of the instrument returns to normal temperature, ask professionals to inspect and eliminate the fault before continuing to use it to prevent accidents.